  • DemoIntegrateTilted

    v5 v6  
    6161  When done, press the right mouse button well away from any points that have been added [on the Mac, if you have a single-button mouse, hold the Control key down and click].
    6262
    63  The calibration is then performed. First the rings are located and an ellipse is optimized for each ring. The center of each ellipse is noted with a blue "+". The derived calibration results are shown in the Image Controls window.
     63 The calibration is then performed. First the rings are located and an ellipse is optimized for each ring. The indexed rings are shown in blue and center of each ellipse is noted with a blue "+"
     64 ([https://subversion.xor.aps.anl.gov/trac/pyGSAS/attachment/wiki/DemoIntegrateTilted/d1w6.png view]).
     65 The derived calibration results are shown in the Image Controls window.
     66 ([https://subversion.xor.aps.anl.gov/trac/pyGSAS/attachment/wiki/DemoIntegrateTilted/d1w7.png view]).
    6467
    65  To see the actual points selected by the program, click on the "show ring picks?" check button. When this is done, it becomes clear that the points selected for the outermost ring are scattered between that and the next ring. This can be repaired by using the Image Operations/Clear Calibration menu item, lowering the area the program uses for searching for points by changing the "pixel search range" to 10 and then repeating the calibration steps (selecting points again on the first ring and then using the right mouse button.) This produces very slightly better calibration constants.
    66 
    67 
     68 To see the actual points selected by the program, click on the "show ring picks?" check button
     69 ([https://subversion.xor.aps.anl.gov/trac/pyGSAS/attachment/wiki/DemoIntegrateTilted/d1w8.png view]).
     70 When this is done, it becomes clear that the points selected for the outermost ring are scattered between that and the next ring. This can be repaired by using the Image Operations/Clear Calibration menu item, lowering the area the program uses for searching for points by changing the "pixel search range" to 10 and then repeating the calibration steps (selecting points again on the first ring and then using the right mouse button)
     71 ([https://subversion.xor.aps.anl.gov/trac/pyGSAS/attachment/wiki/DemoIntegrateTilted/d1w9.png view]).
     72 This produces very slightly better calibration constants
     73 ([https://subversion.xor.aps.anl.gov/trac/pyGSAS/attachment/wiki/DemoIntegrateTilted/d1w10.png view]).